An Oropharyngeal pH Monitoring Device to Evaluate Patients with Chronic Laryngitis

Yuksel ES, Slaughter JC, Mukhtar N, et al. Neurogastroenterology & Motility. 2013;25(5):e315-e323.

OBJECTIVES: Diagnostics for gastroesophageal reflux disease (GERD) are suboptimal because of limited sensitivity. We performed in vitro and in vivo studies to systematically assess the performance characteristics of an oropharyngeal pH probe.

RESULTS: In the liquid phase, at pH 4–5, the oropharyngeal probe had less deviation from the target value than the standard catheter; deviation in the vapor phase was similar (0.4 pH units). Median (interquartile) time to reach equilibrium pH was significantly (P < 0.001) faster with the oropharyngeal than the standard probe. In comparing simultaneous distal esophageal pH characteristics, 96% of recordings with the new and standard probes were in agreement to within `1.0 pH unit; 71% of recordings were in agreement within `0.5 pH units. Patients with chronic laryngitis had significantly higher levels of oropharyngeal acid exposure at pH <4, <5, and <6, in the upright position than patients with GERD or controls (P < .001).

CONCLUSION: Oropharyngeal pH monitoring appears to be more sensitive than traditional pH monitoring in evaluation of patients with extraesophageal reflux. It is a promising tool in evaluation of this difficult group of patients.

 

Topics: pH monitoring, ENT, 2013

Clinical Library

Quick access to relevant clinical news.

Here is our collection of clinical abstracts, clinician and researcher presentations and other relevant clinical news related to Restech's pH system. You will also find various research and insights covering the reflux spectrum from diagnostics to medical and surgical interventions. Stay tuned for updates!